USB and serial port management

Displays
3-6
Proper function of serial or USB touchscreen displays is dependent on the usage of
the correct RS232 and USB ports on the patient monitor.
NOTE
The patient monitoring system only supports the D19KT and D15K displays with
respect to USB.
Refer to the following table when connecting your displays containing either USB or
serial touchscreens.
Display port
Display 1
DVI-I 1
Display 1 cloned
DVI-D 2
Display 2
Display 3
DVI-I 3
Display 3 cloned
Unused USB ports may be connected to other accessories such as keyboard, mouse,
barcode reader, or remote control.
If the D15K or D19KT displays are not used
If you are not using the D15K or D19KT displays, there are no restrictions on
connecting devices to USB ports.
The patient monitor can support up to three independent displays when equipped with
an optional third video card. The optional third video can only be used to support the
iPanel application. The patient monitor also supports two additional cloned displays
on the DVI-I 1 and DVI-I 3 ports. Cloned displays must be connected using a DVI-I
to DVI-D and VGA splitter and must support analog video.
Refer to the "CARESCAPE Monitor B850 Addendum for Device Compatibility" for
a list of compatible devices.
WARNING
Do not use the monitor without manufacturer approved mounting
attached.
CARESCAPE Monitor B850
Number of displays
Touch interface ports
USB 1 or RS232 1
USB 1 or RS232 1
USB 2 or RS232 2
USB 3
No touch interface available