GE CARESCAPE Monitor B850 Technical Manual page 60

Hide thumbs Also See for CARESCAPE Monitor B850:
Table of Contents

Advertisement

2040384-004D
The following sample shows the corresponding entries on the Admit/Discharge
Configuration window.
Configure character delimited parser information
Points to note
If you configure Age, you must either select the Age Unit item or one of the age
units (e.g., Years, Months, Weeks, Days) under Fixed Option.
If you configure Height, you must either select the Height Unit item or one of
the height units (e.g., Feet, Inches, Meters, Centimeters, Millimeters) under
Fixed Option.
If you configure Weight, you must either select the Weight Unit item or one of
the weight units (e.g., Kilograms, Grams, Micrograms, Pounds, Ounces) under
Fixed Option.
For an example of admit/discharge configuration for a length delimited parser,
refer to
"Sample character delimited parser information entry"
1.
In the Position column of the Admit/Discharge Configuration window, enter the
sequence number of each item included in the barcode. Use incremental numbers
from 1 (the left-most field) up to 16 (the right-most field).
If an item is not included in the barcode, leave the Position field blank for the
item.
CARESCAPE Monitor B850
on page 4-18.
4-17