Using A Service Pc Over The Network - GE CARESCAPE Monitor B850 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B850:
Table of Contents

Advertisement

Using a service PC over the network

Using a service PC with a crossover cable
2040384-004D
1.
Verify that the patient monitor is connected to a live IX Network.
2.
Record the IP and Netmask addresses of the patient monitor:
Static IP address: ______________________
Netmask: ________________________
3.
Verify that the service PC is on the same network as the patient monitor by
performing a network ping to the patient monitor. If needed, configure the
service PC's IP and Netmask addresses.
4.
From a computer connected to the same network as the patient monitor, launch a
web browser.
5.
In the Address field, type https://[IX IP address]:10000 and press Enter.
NOTE
[IX IP address] is the IX Network IP address of the patient monitor.
The Login to Webmin dialog box displays.
6.
Continue to
"Log into Webmin"
1.
Verify that the IX Network on the patient monitor is configured in manual
configuration mode.
2.
Record the IP and Netmask addresses of the patient monitor:
Static IP address: ______________________
Netmask: ________________________
3.
Verify that the service PC's network is configured in manual configuration mode
to allow communication with the patient monitor. If needed, configure the
service PC's IP and Netmask addresses.
4.
Verify that the service PC is able to perform a network ping to the patient
monitor. See
"Ping a TCP/IP network device"
5.
Connect a correctly configured service laptop to the CARESCAPE Network IX
connection port on the patient monitor using a crossover cable.
CARESCAPE Monitor B850
on page 4-4.
on page 8-5.