PRN 50-M digital writer

Remote and keypad
Alarms
Parameters
Auxiliary devices
2040384-004D
If present, verify that the PRN-50 digital writer is able to print clearly and all
characters are uniform in size and contrast.
Verify that all keys are functioning properly.
Verify that the audio and visual alarm signals are working properly.
1.
Set a parameter alarm limit outside of the current measured patient values.
2.
Confirm that the following alarm notification events occur:
The audible alarm sounds the correct tone.
The alarm light illuminates, if present.
The value flashes in the parameter block with the correct color.
An alarm graph prints (if enabled).
NOTE
See the "Alarms" section in the user's manual for a description of alarm
behavior.
3.
Return the parameter alarm limit to the original value.
For parameter planned maintenance checkout, refer to the technical manual for the
acquisition module.
For planned maintenance for all other devices connected to USB, serial and M-ports,
refer to the respective technical manual.