Problem
Unable to view historical
12SL data
Unable to view and set
reminder volume
Unable to upload the
software
8-14
Possible cause
12SL is not figured correctly.
The MC Network is down between
CARESCAPE Monitor B850 and MUSE.
The MUSE server is not working.
The national requirement is set to France.
CARESCAPE Monitor B850
Action
Refer to

"MUSE/12SL"

on page 4-13.
Check for MC Network connection.
1. Log onto Webmin > Configuration > National
Requirements.
2. Select None.
See "CARESCAPE Monitor B850 Software Installation
Instructions".