Replace the speaker

2040384-004D
1.
Open the unit as described in steps 1 and 2 on page 9-4.
2.
Disconnect the speaker and power supply cables.
3.
Remove 4 screws holding the power supply to the chassis.
4.
Remove the power supply PCB to access the speaker screws.
5.
Remove 2 screws and washers holding the speaker to the front bezel. Remove the
speaker.
6.
Replace the speaker.
7.
Reassemble the power supply PCB.
8.
Reassemble the patient monitor in the reverse order of disassembly.
9.
"Replace the cover on the unit"
10. Complete the procedures in
CARESCAPE Monitor B850
on page 9-10.
"Recommended checkout"
on page 9-10.