STP/TP/ST settings
P/PT/PP settings
Passwords
4-24
NOTE
This procedure applies to the E-PRESTN, E-PRETN, E-RESTN, E-PSM and E-
PSMP modules (which use SpO2, invasive pressure or temperature
measurements) and is needed after corrective maintenance only. For more
information, refer to the "Module Frames and Modules Technical Manual".
1.
If you have not already done so,
Configuration tab.
2.
On the Configuration tab, select Modules.
3.
On the Sub-Modules for Modules menu, select STP/TP/ST Settings.
4.
Select a new configuration from the New Configuration drop down list.
5.
Select Submit.
The changes will take effect after the patient monitor is restarted.
NOTE
This procedure applies to the E-P, E-PT, and E-PP modules (which use invasive
pressure or temperature measurements) and is needed after corrective
maintenance only. For more information, refer to the "Module Frames and
Modules Technical Manual".
1.
If you have not already done so,
Configuration tab.
2.
On the Configuration tab, select Modules.
3.
On the Sub-Modules for Modules menu, select P/PT/PP Settings.
4.
Select a new configuration from the New Configuration drop down list.
5.
Select Submit.
The changes will take effect after the patient monitor is restarted.
WARNING
Control of this user's password is critical to ensure that Webmin on
this device is accessed by only trained and authorized personnel.
Failure to limit access of Webmin to trained and authorized
personnel only may compromise patient safety and/or system
performance.
1.
If you have not already done so,
Configuration tab.
2.
On the Configuration tab, select Passwords.
CARESCAPE Monitor B850
"Log into Webmin"
on page 4-4 and select the
"Log into Webmin"
on page 4-4 and select the
"Log into Webmin"
on page 4-4 and select the
2040384-004D