2040384-004D
3.
To remove a parameter module, grasp it firmly, press the release lever on bottom
of module, and pull out of the guides.
To install a PDM module
WARNING
Do not use identical measurement modules or modules that map a
measurement to the same channel or parameter window. If such
modules have been connected, remove the module that has been
most recently connected. You can also remove both modules and
re-connect the new module after five seconds.
The PDM module can be connected directly to the patient monitor or it can be
mounted to F5 Frame insertion guides.
NOTE
PDM software v1.2 is required for use with CARESCAPE Monitor B850
software.
The PDM can not be installed to an F7 Frame.
Refer to the "CARESCAPE Monitor B850 Addendum for Device
Compatibility" for a list of compatible devices.
When the PDM is used without a battery, it is necessary to allow additional
time to power up. Do not interrupt the startup sequence by unplugging the
PDM.
To connect the PDM module to the patient monitor – Connect one end of the
ePort cable to the PDM and the other end of the ePort cable to the ePort/Tramnet
connector on the patient monitor.
CARESCAPE Monitor B850
3-17