9-16
6.
Use a wrench to disassemble the rotary switch.
7.
Replace the rotary switch and the main TRIM KNOB.
8.
Reassemble the display in the reverse order of disassembly.
NOTE
The torque specification for tightening the screws and encoder washer/nut is
1.13 + 0.11 Nm (10.0+1.0 lbf-in).
9.
Complete the procedures in
CARESCAPE Monitor B850
"Recommended checkout"
on page 9-20.