Software Reset - GE Definium 5000 Operator's Manual

X-ray system
Table of Contents

Advertisement

Software Reset

Software reset is used to restore the system software. A software reset may take1-2 minutes to
complete.
NOTE: Software reset does not require Shutdown.
NOTE: The system will not be available for acquiring images during the reset cycle.
Follow this process to perform a software reset.
1. Close all open exams.
2. On the Worklist or Image Management screen, press [UTILITIES].
The System – System screen (Figure 4-5) appears.
Figure 4-5 System – System screen with Software Reset button
3. Press [SOFTWARE RESET].
A message appears: "Press OK to restart Software Application. Press Cancel to return."
4. Press [OK].
A "Loading Application" progress bar appears on screen.
The login screen appears once the software has been reset.
5. Login to the system to begin work.
Definium 5000 X-Ray System
Operator Manual
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter 4: General Information
4-6

Advertisement