Communication - GE CARESCAPE Monitor B850 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B850:
Table of Contents

Advertisement

Remote displays
2040384-004D
NOTE
All installations should be compliant with IEC 60601-1-1 and local electrical
codes.
Non isolated communication lines
If complete isolation is not required, this method will provide the most cost effective
means of extending your USB installation. This type of installation should not be
used for connections to non-medically used rooms per IEC 60601-1-1.
Displays may be extended up to 15 meters from the patient monitor using the
following cables:
15-meter DVI-I cable (p/n 2042766-001)
The connections for the DVI-I cable are the same as any other video cable.
5-meter USB extender (p/n 2042768-001)
Type A
receptacle
In the following example, two 5-meter USB extenders, plus a standard 5-meter USB
cable extend the remote display up to 15 meters from the patient monitor. See
instructions"
on page 3-10 for instructions on setting up the remote display as shown
in the example.
NOTE
Only the display interface connections (video and USB connections) are shown.
Not all other connections are shown.
CARESCAPE Monitor B850
Type A plug
"Setup
3-9