Additional SCPI Commands
:CLEar
:STATus:QUEue:CLEar
Description
This command is used to clear the Error Queue of all messages.

:SYSTem subsystem

The SYSTem subsystem commands are summarized in
:PRESet command
:PRESet
:SYSTem:PRESet
Description
This command returns the instrument to states optimized for front panel operation.
:SYSTem:PRESet defaults are listed in the SCPI tables
Performance commands
:FAZero[:STATe] <b>
:SYSTem:FAZero[:STATe] <b> Control Front Autozero
Parameters
<b>= 0 or OFF
Description
With Front Autozero enabled (which is the default setting), the instrument performs two A/D
measurement cycles for each reading. The first one is a normal measurement cycle. The second
measurement cycle is performed with the polarity of the front-end amplifier reversed. This
current-reversal measurement technique is used to cancel internal offsets in the amplifier. With
Front Autozero disabled, the second measurement cycle is not performed.
The speed for Delta measurements can be doubled by disabling Front Autozero. Delta uses
its own current-reversal technique to cancel offsets. Therefore, the dual measurement technique
of Front Autozero is not required.
See
drawbacks) on Front Autozero.
Disable Front Autozero
1 or ON
Enable Front Autozero
"Autozeroing
modes" in
Section 2
Clears all messages from Error Queue
Return to :SYSTem:PRESet defaults
(Table 14-1
for more information (including benefits and
Table
14-9.
through
Table
14-12).