Figure 3-2 Moving And Repeating Filters - Keithley 2182 User Manual

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

3-10
Range, Digits, Rate, and Filter
Figure 3-2
Moving and repeating filters
Conversion #10
Conversion #1
Conversion #10
Conversion #1
Digital filter example
Filter Count = 10
Filter Window = 0.01% of range
Filter Type = Moving
Ten readings fill the stack to yield a filtered reading. Now assume the next reading (which is
th
the 11
be loaded with that same reading. Each subsequent valid reading will then displace one of the
loaded readings in the stack. The FILT annunciator will flash until 10 new readings fill the stack.
NOTE
Conversion #11
#9
#8
#7
#6
Reading
#5
#1
#4
#3
#2
Conversion
A. Type - Moving Average, Readings = 10
Conversion #20
#9
#8
#7
#6
Reading
#5
#1
#4
#3
#2
Conversion
B. Type - Repeating, Readings = 10
) is outside the window. A reading will be processed (displayed); however, the stack will
Bit 8 of the Operation Event Status Register sets when the filter window has properly
settled. See
"Status
structure" in
Conversion #12
#10
#9
#8
#7
Reading
#6
#2
#5
#4
#3
#2
Conversion #3
Conversion #30
#19
#18
#17
#16
Reading
#15
#2
#14
#13
#12
#11
Conversion #21
Section 11
for details.
#11
#10
#9
#8
Reading
#7
#3
#6
#5
#4
#29
#28
#27
#26
Reading
#25
#3
#24
#23
#22

Advertisement

Table of Contents
loading

This manual is also suitable for:

2182a

Table of Contents