Query commands
The Query command requests the presently programmed status. It is identified by the
question mark (?) at the end of the fundamental form of the command. Most commands have a
query form.
:TRIGger:TIMer?Queries the timer interval
Most commands that require a numeric parameter(<n>) can also use the DEFault, MINimum,
and MAXimum parameters for the query form. These query forms are used to determine the
*RST default value and the upper and lower limits for the fundamental command.
:TRIGger:TIMer? DEFault
:TRIGger:TIMer? MINimum
:TRIGger:TIMer? MAXimum
Case sensitivity
Common commands and SCPI commands are not case sensitive. You can use upper or lower
case and any case combination. Examples:
*RST
:DATA?
:SYSTem:PRESet
Long-form and short-form versions
A SCPI command word can be sent in its long-form or short-form version. The command
subsystem tables in
Section 14
indicated by upper case characters.
:SYSTem:PRESet
:SYST:PRES
:SYSTem:PRES
Note that each command word must be in either long-form or short-form. For example,
:SYSTe:PRESe is illegal and will generate an error. The command will not be executed.
Queries the *RST default value
Queries the lowest allowable value
Queries the largest allowable value
=
*rst
=
:data?
=
:system:preset
are in the long-form version. However, the short-form version is
long-form
short form
long-form and short-form combination