Additional SCPI Commands

:BEEPer command

:STATe <b>
:SYSTem:BEEPer:STATe <b>
Parameters
<b> = 1 or ON
Description
This command is used to enable or disable the beeper for limit tests and HOLD.
:KCLick command
:KCLick <b>
:SYSTem:KCLick <b>
Parameters
<b> = 1 or ON
Description
This command is used to enable or disable the keyclick. The keyclick can also be enabled or
disabled from the front panel by pressing SHIFT then LOCAL.
:POSetup <name> command
:POSetup <name>
:SYSTem:POSetup <name>
Parameters
<name> =
Description
This command is used to select the power-on defaults. With RST selected, the instrument
powers up to the *RST default conditions. With PRES selected, the instrument powers up to the
:SYStem:PRESet default conditions. Default conditions are listed in the SCPI tables
(Table 14-1
With the SAV0 parameter selected, the instrument powers-on to the setup that is saved in the
specified location using the *SAV command.
Enable beeper
0 or OFF
Disable beeper
Enable keyclick
0 or OFF
Disable keyclick
RST
Select *RST defaults on power up
PRESet
Select :SYSTem:PRESet defaults on power up
SAV0
Select saved defaults on power up
through
Table
14-12).
Enable or disable beeper
Enable or disable keyclick
Program power-on defaults