Figure; Delta Measurement Connections; Figure 5-3 Delta Measurement Connections - Keithley 2182 User Manual

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

Figure 5-3

Delta measurement connections

2182
CH 1
Step 3
Configure the trigger model of the SourceMeter.
The menu structure to configure triggers is accessed by pressing CONFIG and then TRIG.
Configure the trigger model as follows:
Arm-Layer:
Arm-In Event
Arm-Out TLink Line
Arm-Out Events
Arm Count
Trig-Layer:
Trigger-In:
Trigger-In Source
Trig-In TLink Line
Event Detect Bypass
Trigger In Events
Trigger-Out:
Trig-Out TLink Line
Trigger Out Events
Delay
Trig-Count
8501 Trigger Link Cable
HI
DUT
0.1Ω
LO
IMMEDIATE
#3
OFF
INFINITE
TRIGGER LINK
#1
NEVER
SOURCE=ON DELAY=OFF MEAS=OFF
#2
SOURCE=ON DELAY=OFF MEAS=OFF
000.0000 s
0002 (trigger count must equal the number of sweep points)
Ratio and Delta
SourceMeter
5-11

Advertisement

Table of Contents
loading

This manual is also suitable for:

2182a

Table of Contents