Also note that channel voltage differential reduces the maximum measurement capability of
Channel 2. Normally, Channel 2 can measure up to 12V. However, a 2V differential reduces the
maximum measurement capability of Channel 2 to 10V. In
Channel 2 will cause an overflow condition.
NOTE
Figure 2-5
Connections - dual channel voltage
2107
Input Cable
red
HI
DCV1
CH 1
black
LO
HI
green
CH 2
DCV2
LO
white
2182
A. Typical Measurement Configuration

Temperature only connections

Channel 1 of the Model 2182 can be used to make temperature measurements.
shows connections using the internal reference junction. Keep in mind that the thermocouple
wires must be soldered directly to a LEMO connector as previously explained.
Figure 2-6
Connections - temperature (internal reference)
Thermocouple wire
soldered directly to
LEMO connector
(one of two)
HI
CH 1
LO
HI
CH 2
LO
2182
Channel 2 HI or LO cannot be more than 12V peak from Channel 1 LO.
Cable-to-copper
wire connection
(one of four)
DUT
DUT
Test Circuit
red
DUT
TEMP1
black
Test Circuit
Thermocouple
Voltage and Temperature Measurements
Figure
red
HI
DUT
#1
DCV1
CH 1
black
LO
R
HI
green
CH 2
DCV2
(10V range)
DUT
LO
#2
white
2182
Note: Channel 2 HI or LO must not exceed 12V from Channel 1 LO.
B. Voltage Differential Between Channels
2-15
2-5A, a >10V input to
7V
10V
2V