Voltage and Temperature Measurements
Controlling autozeroing modes
For front panel operation, the two autozeroing modes are controlled from the SHIFT >
CONFIG menu as follows:
NOTE
1.
2.
3.
4.
5.
6.
NOTE

LSYNC (line cycle synchronization)

Synchronizing A/D conversions with the frequency of the power line increases common
mode and normal mode noise rejection. When line cycle synchronization is enabled, the
measurement is initiated at the first positive- or negative-going zero crossing of the power line
cycle after the trigger.
conversions. If the trigger occurs during the positive cycle of the power line (as shown in

Figure

line cycle. If the next trigger (Trigger #2) occurs during the negative cycle, then the
measurement process starts with the positive-going zero crossing.
Figure 2-1

Line cycle synchronization

1 PLC
Trigger
#1
A/D
Conversion
Phase A
For remote programming, the commands to control the two autozeroing modes are
listed in
Table
2-2.
Press SHIFT and then CONFIG to display the present state of Front Autozero; Y = yes
(enabled), N = no (disabled).
To change the FRONT AZERO setting, use the
If you do not wish to view or change the Autozero setting, jump to step 6. Otherwise,
proceed to the next step.
Press the
key to display the present state of Autozero; YES (enabled), NO (disabled).
To change the AUTOZERO setting, use the
Press ENTER to enter the setting(s) and exit from the menu structure.
The factory default setting for Front Autozero and Autozero in ON (enabled). The set-
tings can be saved in the user default setup (see
Figure 2-1
2-1), the first A/D conversion starts with the negative-going zero crossing of the power
Reading
Done
A/D
Conversion
Phase B
or
or
"Default
shows the measurement process that consists of two A/D
Trigger
#2
A/D
Conversion
Phase A
key to display Y or N.
key to display YES or NO.
settings" in
Section
Reading
Done
A/D
Conversion
Phase B
1).