12-12
Common Commands
*RST – Reset
Description
When the *RST command is sent, the Model 2182 performs the following operations:
1.
2.
3.
NOTE:
*SAV – Save
Parameters
<NRf>=0
Description
Use the *SAVE command to save the present instrument setup configuration in memory for
later recall. Any control affected by *RST can be saved by the *SAV command. The *RCL
command is used to restore the instrument to the saved setup configuration.
Only one setup configuration can be saved and recalled.
NOTE:
*SRE <NRf> – Service Request Enable
*SRE? – Service Request Enable Query
Parameters
<NRf>
Return 2182 to *RST defaults
Returns the Model 2182 to the *RST default conditions (see SCPI tables).
Cancels all pending commands.
Cancels response to any previously received *OPC and *OPC? commands.
For RS-232 operation (and in some cases, GPIB operation), *OPC or *OPC? should
be used with *RST, which is a slow responding command. Details on *OPC and
*OPC? are provided in
Save present setup in memory
For RS-232 operation (and in some cases, GPIB operation), *OPC or *OPC? should
be used with *SAV, which is a slow responding command. Details on *OPC and
*OPC? are provided in
=
0
Clears enable register
1
Set MSB bit (Bit 0)
4
Set EAV bit (Bit 2)
8
Set QSB bit (Bit 3)
16
Set MAV bit (Bit 4)
32
Set ESB (Bit 5)
128
Set OSB (Bit 7)
255
Set all bits
Section
12.
Section
12.
Program service request enable register
Read service request enable register