Limits

Setting limit values

Use the following steps to enter high and low limit values:
1.
2.
3.
4.
5.
6.
7.
8.
Enabling limits
Use the following procedure to turn on the limits operation:
1.
2.
3.
4.
Press the Limits VALUE key to view the present HI1 limit value:
HI1:+1.000000 ^
(default)
To change the HI1 limit, use the cursor keys (
( and
) to display the desired value. Move the cursor to the rightmost position (^) and
use the ( and
) keys to move the decimal point. Note that with the cursor on the
polarity sign, pressing
or
Press ENTER to view the present LO1 limit value:
LO1:–1.000000 ^
(default)
Enter the desired value for this low limit.
Press ENTER to view the present HI2 limit value:
HI2:+2.000000 ^
(default)
Enter the desired value for this high limit.
Press ENTER to view the present LO2 limit value:
LO2:–2.000000 ^
(default)
Enter the desired value for this low limit and press ENTER to return to the normal
display.
Press the Limits ON/OFF key to view the present beeper status:
BEEP: NEVER
(default)
To change the beeper setting, use the
INSIDE.
Press ENTER to return to the normal display. The HI/IN/LO status is displayed along
with the reading.
To disable Limits, press the ON/OFF key.
and
) and the manual range keys
toggles the polarity of the value.
and
keys to display NEVER, OUTSIDE, or