Figure 9-4 Waveform To Be Programmed Into Model 2400; Application — I-V Curves Using Internal Scan - Keithley 2182 User Manual

Nanovoltmeter
Hide thumbs Also See for 2182:
Table of Contents

Advertisement

9-14
Stepping and Scanning
Application — I-V curves using internal scan
SCAN for IV curves [Measure V, sweep I, constant H (magnetic field) or
T (temperature)]
SCAN can be used to measure V, while sweeping the current through a sample with a constant
magnetic field or a constant temperature. With the use of a Keithley Model 2400 SourceMeter
and a Model 8501 Trigger Link cable, (see
can be programmed to sweep the current in a bipolar, (DC current-reversal technique) growing
amplitude fashion. See waveform in
can be recalled at the end of the sweep. By having the 2400 and 2182 in a tight hardware control,
the DC current-reversal technique can be run at a rate of 8/sec. at a 1PLC integration time. This
will greatly reduce any thermal EMFs in the system by being able to reverse the DC current
before any temperature effects can occur. Looking for critical currents I
at a faster reading rate.
NOTE
Figure 9-4
Waveform to be programmed into Model 2400
+5mA
+2mA
+1mA
0
–1mA
–2mA
–5mA
Channel 1 is used to measure the voltage across the sample, while Channel 2
measures the voltage across a known reference (R
current in the sample.
P8
P4
P6
P0
P2
P1
P3
P5
P7
Section 7
for more details on triggering), the 2400
Figure
9-4. This sweep will store into the 2182 memory and
REF
P10
...
P9
P11
can be accomplished
C
) resistor to determine the