Programming example
The following program fragment configures the Filter for Channel 2 voltage (DCV2). It
disables the analog filter and enables the digital filter (5% window, count 10, moving).
'Analog Filter:
CALL SEND(7,":sens:volt:chan2:lpas off",status%)

' Digital Filter:

CALL SEND(7,":sens:volt:chan2:dfil:wind 5",status%)
CALL SEND(7,":sens:volt:chan2:dfil:coun 10",status%)
CALL SEND(7,":sens:volt:chan2:dfil:tcon mov",status%)
CALL SEND(7,":sens:volt:chan2:dfil:stat on",status%)
Range, Digits, Rate, and Filter
'Disable analog filter.
'Set window to 5%.
'Set count to 10.
'Select moving
'filter.
'Enable digital
'filter.
3-13