Filter change

1. If the filter is visibly contaminated, it must be replaced
under all circumstances.
2. Replace the filter if the pump performance drops
(audibly).
In most cases, it is sufficient to replace the flue gas filter.
- To replace a filter, remove the filter cover by twisting to the
left.
- Take out the used filter and insert the new filter.
- Screw on the filter cover.
- The cross strut of the filter housing must be aligned with
the markings on the box housing.
Spare filters with Part no. 0554.3381 available.

Condensation trap

To be observed
- Only empty the condensation trap when the pump is
switched off.
- Do not damage the sealing rings when assembling the
condensation trap.
- Pull out the condensation trap horizontally to empty the
condensation.
- Open the drain plug.