Advertisement

CO meas.
range
Low
1
High
2
1
corresponds to: undiluted
2
corresponds to: 20xdilution
CO measuring ranges with CO Low or CO High
CO concentration
in sensor (ppm)
from........to
0
1000
50
1500
Notes:
The given measuring ranges are recommendations to
achieve the highest possible accuracy. Higher deviation is to
be expected outside these measuring ranges.
Note: The toxic sensors may be damaged if CO
concentrations are greater than 40 000 ppm. Remove the
toxic sensors during applications of this kind in order to
prevent the sensors from being damaged. Measurements
can be carried out in this range using the O2 sensor.
CO`s lower explosion limit is reached if CO concentrations of
more than 10.8 % (108 000 ppm) come into contact with air.
The lower explosion limit is significantly lower in combusion
processes such as wood combustion due to the nonburned
hydrocarbons and hydrogen.
Solid fuel measurement option
Measuring range CO
(ppm)
from........to
00
1000
1000
30000
for analyser box
Accuracy
See Technical data
± 7.5 % of read.