Maintenance

Upgrades

Heater measurement cell
Plastic
Hose
adapter
connector
98
1.
Switch off the device.
2.
Open the measuring cell cover.
Upgrade modules are connected at one of the empty
slots. The measuring modules NO and HC must only
be inserted at the slots marked "NO" and "Type A/HC"
– observe switch settings. Connect the measuring
modules NO
"Type A" (see label in the measuring cell cover).
3.
Remove the hose connector and plastic adapter of an
empty slot.
4.
Push on the extension module.
5.
Connect the hose connector to the gas hose nipple of
the extension module.
6.
Install the cell heater on the extension module.
7.
Close the measuring cell cover.
8.
Switch on the device with the control unit or using the
control software.
9.
Insert the added parameter to the display sequence.
, SO
, H
S at any desired slot marked
2
2
2