Differential Pressure Measurement With The Control Unit - TESTO 300 XXL Instruction Manual

Hide thumbs Also See for 300 XXL:
Table of Contents

Advertisement

Example measurement
Differential pressure measurement
with the control unit
Measurement menu
1 sec
Function menu
Select "dP1" and/or
"dP2"
Measurement menu
dP1
Automatic sensor
zeroing
10 sec
Start measurement
01:00 min
T T S S t t a a r r t t
26
Preparing for measurement
Selection of a suitable measuring range.
Two measuring ranges P1 (-80 to +80 hPa) and P2 (-1000
to +1000 hPa) are available in the function menu:
Start measurement
1 Device selection: control unit
2 Activate the function menu with the
the function keys. Select the function key
range) or
dP2
confirm with
3 Zero sensor with function key
4 Start measurement with function key
measurement with function
ment is terminated automatically when the time has
expired (Input menu - Timer).
5 Display and printing of the current reading.
Do not use in the main duct!
Ensure that there are no leaks between the sampling
point and the device (escaping gas and atmospheric
oxygen can create an explosive mixture)!
Do not smoke or use naked flames during the
measurement!
Observe the rated pressure of the connections!
Do not exceed the measuring range!
No gas enters the device because an air cushion forms
between the sensor and the gas inlet. The measuring unit
is therefore not protected against explosions.
Stop measurement
00:27 min
T T S S t t o o p p
(high range) with
.
dP1
TStop
Print
key and one of
(low
dP1
and
.
, stop
TStart
key or the measure-
Measurement menu