Change Co, Co H2 And No Sensor - TESTO 300 Pro Smoke Edition Manual

Combustion analyzer
Table of Contents

Advertisement

Find Quality Products Online at:
10 Maintenance

10.7.2 Change CO, CO H2 and NO sensor

Measuring instrument is open, see Section Open measuring
instrument.
1
Remove faulty sensor and hose
connections from the slot.
2
Remove hose connections from the
faulty sensor/bridge.
For NO sensor: Remove auxiliary circuit board.
Do not remove the auxiliary circuit board of the NO sensor until
immediately before installation. Do not leave the sensor without
auxiliary circuit board for longer than 15 minutes.
3
Fit the hose connections onto the new sensor.
4
Fit new sensor into the slot and at the same time fit the hose
connections onto the gas path connections.
Make sure that the socket on the sensor circuit board is
correctly connected to the contact plug.
5
Close the measuring instrument.
68
GlobalTestSupply
www.
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

300 pro300 smoke edition300Pro

Table of Contents