Table 15 Pinout for the alarm and I/O connector

Pin
I/O
Signal
1
Out
Al1
2
Out
Al2
3
Out
Al3
4
Out
AOUT1
5
Out
AOUT2
6
GND
7
Out
DOUT4
8
Out
DOUT3
9
In
DIN4/
ExtPace
DMTA-20015-01EN [U8778402], Rev. J, January 2014
Description
Alarm output 1. Disabled
on reset, and is at 0 V.
When active, it is at 5 V.
Alarm output 2. Disabled
on reset, and is at 0 V.
When active, it is at 5 V.
Alarm output 3. Disabled
on reset, and is at 0 V.
When active, it is at 5 V.
Analog output 1
Analog output 2
Ground
Digital output 4
Digital output 3
Digital input 4/external
pace input
Programmable input. Can
be configured as generic
input 4, or as an external
pace input (high level
with a minimum signal
length of 50 ms when
used as DIN4, or 21 µs as
ExtPace).
Refer to the
OmniScan MXU Software
User's Manual
("Configuring the Digital
Input" section) for more
information on
programming this input.
Current
±15 mA
±15 mA
±15 mA
±10 mA
±10 mA
N/A
±15 mA
±15 mA
N/A
Connector References
Level
TTL
TTL
TTL
±5 V
±5 V
N/A
TTL
TTL
TTL