Table 13 Pinout For The Scanner Interface Lemo Connector - Olympus OmniScan MX2 User Manual

Hide thumbs Also See for OmniScan MX2:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. J, January 2014

Table 13 Pinout for the scanner interface LEMO connector

Pin
I/O
1
In
2
Out
3
In
4
In
64
Chapter 6
Signal
AIN
Analog input
+5 V
External power supply
This power is shared with pin 2
of the serial connector (see
section 6.1 on page 62).
DIN1/
Digital input 1/preset axis 1
Preset1
Programmable input. Can be
configured as generic input 1, or
as the preset of encoder 1. Refer
to the OmniScan MXU Software
User's Manual ("Configuring the
Digital Input" section) for
information on programming
this input.
To preset, you must use a high-
level signal with a minimum
signal length of 50 ms.
DIN2/
Digital input 2/preset axis 2
Preset2
Programmable input. Can be
configured as generic input 2, or
as the preset of encoder 2. Refer
to the OmniScan MXU Software
User's Manual ("Configuring the
Digital Input" section) for
information on programming
this input.
To preset, you must use a high-
level signal with a minimum
signal length of 50 ms.
Description
Current
Level
±2.5 mA
±5 V
(2 kΩ
input
impedan
ce)
500 mA
N/A
N/A
TTL
N/A
TTL