Possible solutions
Make sure that the instrument is equipped with the video input option.
Make sure that the video cable is correctly connected.

10.8 Video Output Problems

The video output does not function.
Possible solutions
Make sure that the video output option, which must be purchased separately, is
available.
Make sure that the video cable is connected correctly.
Make sure that the external video monitor can support a resolution of
800 × 600 pixels in VGA mode.
Make sure that your monitor supports the VGA standard. The VGA standard
requires the removal of pin 9 from the video connector. While the OmniScan MX
meets this requirement, some monitors do not. If pin 9 has not been removed
from your monitor connector, you will need to remove it. Olympus is not
responsible for any potential damage that may result from pin removal.

10.9 Networking Problems

Ethernet communication is unavailable.
Possible solutions
Make sure that the Ethernet option, which must be purchased separately, is
available.
Make sure that the Ethernet cable is connected correctly.
Make sure that you have installed the software that is required for
communication between the PC and the OmniScan MX.

10.10 USB Peripheral Problems

Some USB peripherals do not function when connected to the OmniScan MX.
DMTA-20015-01EN [U8778402], Rev. J, January 2014
Troubleshooting
107