Acoustic Specifications; Table 35 Acoustic Specifications; C.1 Acoustic Specifications - Olympus OmniScan MX2 User Manual

Hide thumbs Also See for OmniScan MX2:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. J, January 2014
C.1

Acoustic Specifications

This section describes the acoustic specifications for the pulser, receiver, and beam
formation (see Table 35 on page 136).
Voltage
Pulse width
Fall time
Pulse shape
Output impedance
Gain range
Input impedance
System bandwidth
Scan type
Aperture
Number of elements
Number of focal laws
Delay range transmission 0 µs to 10 µs in 2.5 ns increments
Delay range reception
136
Appendix C

Table 35 Acoustic specifications

Pulser
Low: 45 V
High: 90 V
Adjustable from 30 ns to 500 ns; resolution
of 2.5 ns.
<5 ns
Negative square pulse
<30 Ω
Receiver
0 dB to 74 dB in steps of 0.1 dB, maximum
input signal 1.25 Vp-p
50 Ω
0.53 MHz to 21 MHz, ±10 % (–3 dB)
Beam formation
Sectorial and linear
32 elements
128 elements
256
0 µs to 10 µs in 2.5 ns increments

Hide quick links: