Use the ferrite clamp filter with the corresponding cable diameter. The filter must
not slip easily or be difficult to clip onto the cable.
Make sure that the two ferrite clamp-filter parts are closed tight until the clamp
clicks.
1.
Attach the ferrite clamp filter on the cable of the Olympus UT probes, in close
proximity to the connector for the OmniScan MX2.
2.
Attach the ferrite clamp filter on the cable of the Olympus PA probe, in close
proximity to the connector for the OmniScan MX2.
3.
Attach the ferrite clamp filter to the USB cable, in close proximity to the connector
for the OmniScan MX2.
4.
Attach the ferrite clamp filter to the Ethernet cable, in close proximity to the
connector (RJ-45) that connects to the OmniScan MX2.
5.
Attach the ferrite clamp filter to the scanner interface cable, in close proximity to
the connector (LEMO) for the OmniScan MX2.
6.
Attach the ferrite clamp filter to the alarm and I/O cable, in close proximity to the
connector (DE-9) for the OmniScan MX2.
Figure 2-7 on page 43 and Figure 2-8 on page 44 show the applicable connection for
various cables on the OmniScan MX2, and the locations where the ferrite clamp filters
must be attached.
Figure 2-7 Attaching a ferrite clamp filter to a cable (example shown with the
DMTA-20015-01EN [U8778402], Rev. J, January 2014
scanner interface cable)
Basic Operation
43