Data Specifications; Table 73 Data Specifications; N.3 Data Specifications - Olympus OmniScan MX2 User Manual

Hide thumbs Also See for OmniScan MX2:
Table of Contents

Advertisement

On internal clock
External pace
On encoder
N.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 73 on page 191).
Number of data points
Real-time averaging
Rectification
Filtering
Video filtering
Number of points
Range
Maximum slope
A-scan recording
C-scan recording
Maximum file size
Table 72 Acquisition specifications (continued)
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 to 65536 steps

Table 73 Data specifications

Up to 8192
PA: 2, 4, 8, 16
UT: 2, 4, 8, 16, 32, 64
RF, full wave, half wave+, half wave–
PA channel: 3 low-pass,6 band-pass, and 4 high-
pass filters
UT channel: 3 low-pass, 6 band-pass, and 4 high-
pass filters (3 low-pass filters when configured in
TOFD)
Smoothing (adjusted to the probe frequency
range)
Programmable TCG
32: One TCG (time-corrected gain) curve per focal
law
PA: 40 dB
UT:
40 dB/10 ns
6000 A-scans per second (512-point, 8-bit A-scan)
I, A, B, up to 10 kHz
Limited to available internal flash memory:
300 MB
OMNI-M2-PA1664/16128/32128/32128PR Module Specifications
DMTA-20015-01EN [U8778402], Rev. J, January 2014
Processing
per step of 0.1 dB
100 dB per step of 0.1 dB
Storage
191

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omniscan mx

Table of Contents