O.2

Acquisition Specifications

This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 75 on page 196).
.
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
External pace
On encoder
O.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 76 on page 196).
Number of data points
Real-time averaging
196
Appendix O

Table 75 Acquisition specifications

Frequency
100 MHz
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
Display
A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 to 65536 steps

Table 76 Data specifications

Processing
Up to 8192
2, 4, 8, 16, 32, 64