Data Specifications; Table 46 Data Specifications; F.3 Data Specifications - Olympus OmniScan MX2 User Manual

Hide thumbs Also See for OmniScan MX2:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. J, January 2014
F.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 46 on page 150).
Number of data points
Real-time averaging
Rectification
Filtering
Video filtering
Number of points
A-scan recording
C-scan recording
Maximum file size
150
Appendix F

Table 46 Data specifications

Processing
Up to 8000
2, 4, 8, and 16
RF, full wave, half wave+, half wave–
Low-pass (adjusted to probe frequency),
digital filtering (bandwidth, frequency
range)
Smoothing (adjusted to the probe frequency
range)
Programmable TCG
16: one TCG (time-corrected gain) curve per
channel for focal laws
Storage
6000 A-scans per second (512-point, 8-bit
A-scan)
I, A, B, up to 10 kHz
Limited to available internal flash memory:
OmniScan MX: 180 MB (or 300 MB
optional)
OmniScan MX2: 300 MB