Acquisition Specifications; Data Specifications; Table 48 Acquisition Specifications; Table 49 Data Specifications - Olympus OmniScan MX2 User Manual

Hide thumbs Also See for OmniScan MX2:
Table of Contents

Advertisement

G.2

Acquisition Specifications

This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 48 on page 153).
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
External pace
On encoder
G.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 49 on page 153).
Number of data points
Real-time averaging
Rectification

Table 48 Acquisition specifications

Frequency
100 MHz (10 bits)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
Display
A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 to 65536 steps

Table 49 Data specifications

Processing
Up to 8000
2, 4, 8, and 16
RF, full wave, half wave+, half wave–
OMNI-M-PA16128 Module Specifications
DMTA-20015-01EN [U8778402], Rev. J, January 2014
153

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omniscan mx

Table of Contents