System bandwidth
Scan type
Aperture
Number of elements
Number of focal laws
Delay range transmission 0 µs to 10 µs in 2.5 ns increments
Delay range reception
E.2
Acquisition Specifications
This section details the acquisition specifications of frequency, data display, and
synchronization (see Table 42 on page 145).
Digitizing frequency
Maximum pulsing rate
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
OMNI-M-PA1616M and OMNI-M-PA1664M Module Specifications
Table 41 Acoustic specifications (continued)
0.75 MHz to 18 MHz (–3 dB)
Beam formation
Sectorial and linear
16 elements
16 elements for the OMNI-M-PA1616M
64 elements for the OMNI-M-PA1664M
128
0 µs to 10 µs in 2.5 ns increments
Table 42 Acquisition specifications
Frequency
100 MHz (10 bits)
Up to 10 kHz (C-scan)
Display
A-scan: 60 Hz; S-scan: up to 40 Hz
Yes: Volume-corrected S-scan (up to 40 Hz)
Synchronization
1 Hz to 10 kHz
DMTA-20015-01EN [U8778402], Rev. J, January 2014
145