4 Troubleshooting

4.6 Main board and Power Supply

4.6

Main board and Power Supply

Warning messages
Warning code
Description
011
System over-
temperature
Error messages
Error code
Description
001
EPROM error
002
MPWAtemper-
ature
003
Power supply
monitor error
100
Possible cause
Instrument temperature is out
of range (4°C to 35°C)
Possible cause
Instrument temperature is out
of range (4°C to 35°C)
Voltage out of operating range.
Action
Tap Continue on the Touch
screen.
If the system is in a cold
chamber, make sure that the
temperature is not below
+4°C.
Make sure that there is ade-
quate ventilation surround-
ing the system.
Make sure that the system
is not standing in direct sun
light.
System operation is not af-
fected by this warning, the
system can continue to be
used.
Action
Contact a GE Service Engi-
neer.
Contact a GE Service Engi-
neer.
The system cannot be used
in this condition.
Contact a GE Service Engi-
neer.
With assistance from a GE
Service Engineer, check the
Power supply for voltages.