GE AKTA start Maintenance Manual page 107

Hide thumbs Also See for AKTA start:
Table of Contents

Advertisement

Troubleshooting Flow chart 6
1. Switch off the instrument.
2. Disassemble the Conductivity Monitor.
3. Secure the Cable properly and reassemble
the Conductivity Monitor.
Switch on the instrument.
ÄKTA start Maintenance Manual 29060308 AD
ERROR 201
Conductivity Monitor Failure
Error message still
displayed?
No
Proceed with your work.
4.8 Troubleshooting flow charts
1. Replace the Conductivity Monitor.
Yes
2. Enter the new Cell Constant value.
3. Perform Conductivity calibration.
Yes
Has calibration
No
Contact GE Service Engineer.
4 Troubleshooting
passed?
107