Pressure Sensor - GE AKTA start Maintenance Manual

Hide thumbs Also See for AKTA start:
Table of Contents

Advertisement

4.5

Pressure sensor

Error messages
Error code
Description
501
Overpressure
502
Pressure sen-
sor failure
ÄKTA start Maintenance Manual 29060308 AD
Possible cause
Blockage in the tubing,
valves or in the column.
Improperly selected Flow
rate.
Wrong pressure setting used
for chosen application.
The sensor is not connected or
not calibrated.
4 Troubleshooting

4.5 Pressure sensor

Action
Check tubing and valves by
disconnecting one at a time
starting from the Fraction
collector going backwards
towards the Pump. When
pressure is released the
blockage has been found,
clean or replace tube to re-
solve the issue.
Clean the column/separa-
tion media with suitable solu-
tion (1 M NaOH) or replace
the column with a new one.
Check the specification and
pressure of the columns for
correct flow rate.
Perform Zero offset.
Contact a GE Service Engi-
neer.
99