ÄKTA start Maintenance Manual 29060308 AD
Action
Use the up/down arrows to set the desired Flow Rate.
Tap Start flow to start the Pump.
Collect water for at least one minute in the pre-weighed collection tube.
Tap Stop flow to turn the Pump OFF.
Weigh the pre-weighed collection tube containing the collected water
and calculate the volume of the pumped water, then set the Collected
Volume value equal to this volume.
Tap Calibrate.
Result: The following screen opens.
Wait while the calibration is running. The calibration is done when the Cali-
bration screen re-opens.