3.5.4
Reset number of run hours
Instructions
Follow the instructions below to reset the number of run hours of UV Monitor. Reset the
number of run hours when the UV Monitor is replaced.
Step
1
2
3
ÄKTA start Maintenance Manual 29060308 AD
Action
Tap Configuration.
Result: The Configuration screen opens.
Do not change the C drft and C amb values. These parameters must be set
by a GE Service Engineer.
Tap Reset.
Result: A confirmation screen opens.
If the UV Monitor has been replaced, tap Yes.
Or,
If the UV Monitor has not been replaced, tap No to cancel the action.
3 Maintenance operations
3.5 UV Monitor
3.5.4 Reset number of run hours
47