5 Removing and replacing modules

5.8 Injection valve

5.8

Injection valve

Introduction
Replace the complete valve when internal valve parts are damaged, worn or blocked
with salt deposits.
Use the Injection valve kit (see Section 5.9 Injection valve kit, on page 139) to service the
valve when the external parts and liquid distribution parts are damaged, worn or blocked
with salt deposits.
Required tools
Tool
Torx driver
Instructions
Follow the instructions below to remove and replace the Injection valve.
Step
1
136
WARNING
Disconnect power. Always switch off power to the instrument be-
fore replacing any component on the instrument or cleaning the
instrument, unless stated otherwise in the user documentation.
Action
Switch off the instrument by pressing the Power Switch to the O position.
Dimension
T10