GE AKTA start Maintenance Manual page 103

Hide thumbs Also See for AKTA start:
Table of Contents

Advertisement

Troubleshooting Flow chart 2
1. Tighten Flow cell and flush it clean with
DM water.
2. Tighten connectors and remove any
strong direct light on the UV Monitor.
3. Tap Continue .
4. Go to Settings and service screen and
perform UV LED calibration.
ÄKTA start Maintenance Manual 29060308 AD
ERROR 102 UV Monitor Failure
Switch instrument Off, then On again.
Error message
still displayed?
Perform UV LED
Proceed with your work.
Note:
This Error message is displayed only
during switching on, as part of the power
on self test.
1. Tighten and flush the Flow cell.
2. Clean the Flow cell; refer to
Yes
Operating Instructions.
3. Tap Continue on the error display.
Switch instrument Off, then On again.
No
Yes
1. Replace the Flow cell. Follow the Flow cell
replacement procedure.
2. Tap Continue on the error display.
Switch instrument Off, then On again.
No
Yes
1. Switch instrument Off.
2. Disassemble the UV Monitor.
3. Replace the UV Monitor and Flow cell.
4. Switch instrument On.
No
calibration
Contact GE Service Engineer.
4 Troubleshooting
4.8 Troubleshooting flow charts
ÄKTA start
Error message
still displayed?
Yes
Error message
still displayed?
Error message
still displayed?
Yes
103