Delay Volume Setting - GE AKTA start Maintenance Manual

Hide thumbs Also See for AKTA start:
Table of Contents

Advertisement

3.14.3

Delay volume setting

Instructions
Follow the instructions below to set the delay volume.
Step
1
2
Note:
ÄKTA start Maintenance Manual 29060308 AD
Action
In the System screen, tap Delay volume setting.
Result: The following screen opens.
Enter the internal diameter (ID) and length of the tubing from the Outlet
valve to Frac30 in the respective fields, and then tap Save.
Result: The total delay volume from the UV Monitor to Frac30 is displayed
in the Total Delay volume field.
Note:
The delay volume from the UV Monitor to the Outlet valve is constant (0.27 ml)
in all ÄKTA start instruments.
Delay volume setting needs to be set when the tube between Outlet valve
and Frac30 has been replaced.
3 Maintenance operations
3.14 System
3.14.3 Delay volume setting
89