7.4
Replace the main fuses
Required tools
Tool
Flat screwdriver
Procedure
Follow the instructions below to replace the fuses.
Note:
Step
1
ÄKTA go Operating Instructions 29360951 AC
WARNING
Disconnect power. Always disconnect power from the instrument
before replacing fuses.
WARNING
For continued protection from fire hazard, replace only with same
type and rating of fuse.
WARNING
If a fuse requires repeated replacement, do not continue to use
the instrument. Contact an authorized service engineer.
There are two identical fuses at the mains power inlet. Replace both fuses
even if only one is blown.
Action
Disconnect the power cord from the power inlet.
7.4 Replace the main fuses
Dimension
2 to 3 mm
7 Maintenance
97