3 Maintenance operations
3.5 UV Monitor
3.5.3 Flow cell path length
Step
7
46
Action
Update the calculated Cell length value by using up/down buttons or enter
the value. Tap Save path length to save the value to permanent memory.
Note:
The Abs value should be 340 ± 5% mAU, confirming that normalization has
been done.
Note:
For higher precision, instead of using acetone, use the UV test kit, product
number 18112963.
ÄKTA start Maintenance Manual 29060308 AD