GE AKTA start Maintenance Manual page 126

Hide thumbs Also See for AKTA start:
Table of Contents

Advertisement

5 Removing and replacing modules
5.5 Pump
Step
3
126
Action
Remove the two screws from the Pump using a T20 screwdriver, supplied
with the equipment at delivery.
Note:
Make sure that the pump connector is not moved inside the cabinet. The
connector needs to be retained outside the cabinet for re-assembling of a
new Pump module.
Note:
Do not allow the screws to fall into the pump head.