3 Maintenance operations
3.11 Conductivity Monitor

3.11.4 Set reference temperature

3.11.4
Set reference temperature
Description
For the system to calculate the conductivity correctly at different temperatures, the fol-
lowing conditions must be met:
Instructions
Follow the instructions below to set the reference temperature.
Step
1
2
3
74
Enable Temperature Compensation must be checked in the Calibration screen (see
Section 3.11.3 Conductivity flow cell calibration, on page 71).
The temperature at which calibration is performed must be set as a reference tem-
perature.
Action
Go to the Settings and Service screen 3:Conductivity:Configuration screen.
Result: The Configuration screen appears.
Set the Set ref temp value to the temperature noted from the Calibration
screen (see Section 3.11.3 Conductivity flow cell calibration, on page 71).
Note:
The cell constant value is automatically calculated when calibration is per-
formed. Do not change this value.
Tap Save to save the new reference temperature.
ÄKTA start Maintenance Manual 29060308 AD