5 Removing and replacing modules
5.3 UV
Step
6
7
120
Action
Remove the UV Monitor and leave the RJ45 connector as it is.
Connect the RJ45 connector to the new UV monitor.
Note:
Make sure that there is a click sound when the RJ45 connector is connected.
The UV monitor may not work properly if the RJ45 connector is not properly
connected.
ÄKTA start Maintenance Manual 29060308 AD